An Open-Source da Vinci X/Xi Instrument Module for Laparoscopic Surgery Research

Douglas McDonough, Justin Liao, Jiawei Ge, Antony Goldenberg, Yvonne Zhang, Mackenzie A Eagleson, Ravi Prakash, Axel Krieger, Chunwoo Kim
In 30 seconds

An arXiv preprint describes an open-source module that mounts da Vinci X/Xi surgical instruments on a UR5/UR5e arm and gives foot pedals priority over software energy commands. On the bench, it logged all 12 on and off events from six pedal presses. In ex vivo demonstrations, the module recorded energy commands alongside robot motion and video, giving researchers a way to label when cutting and coagulation were requested.

Research question

Can an off-the-shelf robot arm drive da Vinci X/Xi instruments while giving physical pedals priority and recording whether each electrosurgical activation came from a pedal or software?

Problem

Laparoscopic electrosurgery research requires access to instrument motion, energy commands and the timing of both. Clinical platforms restrict low-level access, and standard research adapters generally drive the instrument without exposing who requested energy and when.

Previous approach

MOPS and DASH mount articulated da Vinci EndoWrist tools on commercial arms. Their reported designs cover standard instruments such as graspers and scissors. CRCD and MiDAS record pedal use, and earlier dVRK work triggered energy from software. Those studies did not document both source labels and surgeon priority for software-controlled activation.

New approach

Five DYNAMIXEL XC330-M288-T servos drive spring-loaded discs through timing belts. Instrument-specific coupling matrices map motor movement to joint angles, and motor current helps find zero on joints with detectable mechanical stops. The same driver fits the SureForm 45 Stapler for shaft, wrist and jaw motion. The ESU-TAP v2 interposer uses ROS 2, labels pedal, software and takeover events, and routes pedals directly to the generator when unpowered. CAD, firmware and driver software are available in ARTS.

Results

In bench tests with a DRE ASG-300 set to 10 W and no instrument connected, the board logged 12 of 12 events from six pedal presses. Software timeout was 503.1 ± 3.8 ms over eight trials against a 500 ms setting. The board reported an 11 ms gap during a switch from software CUT to pedal-controlled bipolar. On physical hardware, tool exchange averaged 2.41 ± 0.34 s over 20 trials, excluding homing. That was 0.98 s below the published DASH Xi mean of 3.39 ± 0.51 s. Optical tracking of the Fenestrated Bipolar Forceps put the measured-to-commanded pitch ratio at 0.904 before calibration refinement and 1.00 after it. Yaw moved from 1.14 to 1.02. The Permanent Cautery Hook averaged 34.9 ± 0.2 s to home over ten trials, with joint-equivalent standard deviations of 0.24° to 0.36°. Three ex vivo ARTS tasks demonstrated monopolar and bipolar coagulation at splenic wounds and monopolar cutting during gallbladder dissection.

Limitations

Tissue evidence comes from three qualitative ex vivo demonstrations. The timing measurements came from board telemetry, which reports relay commands. Contact timing and RF delivery were not measured independently. A hung controller or welded relay requires intervention at the generator. Instrument profiles are selected manually, cable-drive hysteresis remains, and staple firing is still under development.

Industry impact

Surgical robot and instrument developers could use the module now to collect labeled bench and ex vivo demonstrations, after checking footswitch wiring and signal levels for their generator. Clinical use would first require independent measurements of delivered energy, fault testing and clinical validation.