ColoACT: Multi-Cue Action Chunking for Smooth Autonomous Colon Navigation on a Self-Propelled Endoscopic Robot
This is a preprint describing ColoACT, an imitation-learning navigation system for a compact self-propelled endoscopic robot. The policy fuses RGB, relative depth, and a gradient-based pseudo-elevation map with an Action Chunking Transformer and temporal ensembling. In ex-vivo porcine colons it reached 85.4% success on straight and 72.5% on curved segments, with smooth steering and success on 90-degree turns (70%) and double bends (60%), suggesting a path toward reducing physician burden in colonoscopy.
Can a multi-cue RGB-depth-elevation observation combined with action chunking and temporal ensembling enable smooth, robust autonomous navigation of a self-propelled colonoscopic robot through ex-vivo porcine colons, including tortuous bends?
Autonomous colon navigation must operate in deformable tissue with weak texture, specular highlights, and viscoelastic contact forces. Geometry-driven pipelines are brittle because they rely on handcrafted features and switching logic, while learned policies can be temporally inconsistent or jittery and suffer from sim-to-real gaps when trained in simulation.
Prior work includes geometry-driven lumen-center fitting and haustral fold detection, reinforcement learning trained in simulated colon models, monocular depth inference, and naive behavior cloning. Geometry methods fail under noisy visuals and non-ideal lumen shapes; simulation-trained RL transfers poorly due to unmodeled viscoelastic contact; step-wise behavior cloning produces high-frequency jitter and short-horizon errors.
ColoACT combines a custom 24×20×34 mm self-propelled Bevel-Gear-Based Endoscopic Robot (BGER) with PDMS treads and a bevel-gear transmission, and a CVAE-based Action Chunking Transformer. The policy takes 320×240 inputs with five channels: RGB, Depth Anything V2-Base relative depth, and a Scharr-gradient log-normalized pseudo-elevation map. A ResNet-18 backbone encodes observations; the decoder predicts 16-step action chunks of normalized linear and angular velocity, and overlapping predictions are fused with a boxcar temporal ensemble. Training used 214 teleoperated ex-vivo porcine trajectories.
On four unseen porcine colons (~60 cm), RGB-D-E ACT achieved 85.4% success on 48 straight trials and 72.5% on 40 curved trials, outperforming RGB-only ACT (35.4%/25.0%), RGB-D ACT (58.3%/50.0%), and RGB-D-E behavior cloning (45.8%/57.5%). In complex segments it succeeded in 70% of 90-degree turns (7/10), 60% of double bends (6/10), and 30% of triple bends (3/10). The ACT policy reduced normalized angular jerk from about 60 a.u. for behavior cloning to below 5 a.u. Offline ADE/FDE over 43 validation samples were 0.0835/0.1689 for RGB-D-E, 23.8% better ADE and 18.6% better FDE than RGB-D. Perception stability over 50 static frames showed depth std 0.0071 and elevation std 0.0250. Real-time control ran at 15 Hz on an RTX 4060.
The authors note the lack of force sensing and validation only in ex-vivo tissue; they plan in-vivo trials to address peristalsis and lubrication. Trial counts are modest (up to 48 per condition, 10 per complex bend), all tests use a single custom robot, and no code or data release is reported. Generalization to living human anatomy and clinical safety remain unproven.
This work is most relevant to medical-device companies building robotic colonoscopy or autonomous endoscopic platforms, including flexible robotic endoscope startups and established endoscopy firms. Research groups can use the policy now; preclinical and product prototyping may follow in 1–3 years if in-vivo and force-sensing results are positive. Clinical deployment is 3+ years away and would require larger animal/human trials, tissue-contact safety mechanisms, regulatory clearance, and evidence of reduced operator workload.